HI9812-5

Portable pH/EC/TDS/Temperature Meter – HI9812-5

Description

The HI9812-5 is an economical, wateproof, portable meter that measures pH, conductivity (EC), total dissolved solids (TDS) and temperature (°C) with a single probe. The meter operation is simplified with dedicated buttons for measurement modes to be be displayed on the LCD and knobs for single point calibration of pH and EC/TDS. The HI9812-5 is most suitable for agricultural, greenhouse and hydroponic applications.

•  3 in 1 probe with pH, EC/TDS and temperature sensors
•  Amplified pH electrode for immunity to electrical noise including from ballasts and pumps
•  Large LCD with instructions

The HI9812-5 portable meter features a large LCD which displays either pH, EC, TDS or temperature readings along with tutorial instructions. The HI9812-5 is a simplistic pH/EC/TDS meter that is calibrated manually to a single point with the use of two trimmers. pH is calibrated to pH 7.01 while EC/TDS is calibrated to either 1413 µS/cm or 1382 ppm. The LCD screen has battery life indicator as well as on-screen tutorial messages.

The supplied HI1285-5 probe is a polypropylene body, amplified pH electrode with a built-in EC/TDS and temperature sensors. The amplifier for the pH electrode prevents interference from humidity and electrical noise from common sources including from motors, ballasts or pumps. The HI9812-5 is a versatile meter best suited for agricultural, greenhouse and hydroponic applications.

  Features at-a-glance
•  All readings are compensated for variations in temperature
•  Automatic Temperature Compensation
•  Factor automatically adjusts from 0.56 to 0.78 based on actual EC readings
•  Factor based on 442 curve for natural water
•  Low Battery Indicator
•  Variable EC to TDS conversion factor
•  Water Resistant

  SPECIFICATIONS

 pH Range : 0 to 14 pH
 pH Resolution : 0.1 pH
 pH Accuracy (@25ºC/77ºF) : ±0.1 pH
 pH Calibration : manual, one-point
 EC Range : 0.00 to 1990 µS/cm
 EC Resolution : 10 µS/cm
 TDS Range : 0 to 1990 ppm (mg/L)
 TDS Resolution : 10 ppm (mg/L)
 EC/TDS Accuracy : ±2% F.S.
 TDS Conversion Factor : 0.5 ppm (mg/L) = 1 µS/cm
 Temperature Range : 0.0 to 60.0°C
 Temperature Resolution : 0.1°C
 Temperature Accuracy (@ 25°C/77°F) : ±0.5°C
 Temperature Compensation : automatic, 0 to 50°C
 Temperature Coefficient : β = 2% /°C
 Probe : HI1285-5 polypropylene body, pre-amplified multiparameter probe with internal temperature sensor, 8-pin DIN connector and 1m cable
 Calibration : manual, one-point
 Battery Type / Life : 9V / approximately 150 hours of continuous use
 Environment  : 0 to 50°C (32 to 122°F); RH max 100%
 Dimensions : 145 x 80 x 36 mm (5.7 x 3.1 x 1.4”)
 Weight : 230 g (8.1 oz.)
 Ordering Information : HI9812-5 is supplied with HI1285-5 multiparameter probe, HI70007 pH 7.01 buffer solution sachet,

HI70442 1500 ppm calibration solution sachet,HI70031 1413 µS/cm calibration solution sachet,

HI700661 electrode cleaning solution sachets (2), 9V battery, instructions and rugged carrying case.

 Warranty : 2 years (probe 6 months)