Nickel Low Range Portable Photometer
HI96740
Description
The HI96740 Nickel Low Range Portable Photometer combines accuracy and ease of use in an ergonomic, portable design. A user can accurately determine the concentration of nickel within a 0.000 to 1.000 mg/L (ppm) range using the HI93740-0 ready made reagents. The HI96740 offers many advanced features including the exclusive CAL Check™ function used for performance verification and calibration of the meter.
• Automatic Shut Off
• Built-in Timer
• CAL Check™ Validation and Calibration
The HI96740 portable photometer is for the low range measurement of nickel. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check™ feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.
Features/Benefits:
Auto-shut off
• Automatic shut off after 10 minutes of non-use when the meter is in measurement mode.
• Prevents wastage of batteries in the event the meter is accidentally not turned off.
Built in timer
• Ensures the appropriate reaction time is used for the chemical reaction. At the end of the timer the meter will automatically take the reading. This feature ensures consistency across multiple users.
CAL Check™
• Allows for performance verification and calibration of the meter using a CAL Check NIST traceable secondary standard.
Cooling Lamp Indicator
• It is necessary to maintain a consistent temperature of the optical components in order to maintain a narrow wavelength band of light.
• This photometer has a cooling lamp indicator that is displayed for a short period of time before each measurement in order to allow the components to cool and obtain the highest accuracy possible.
Error messages
• Messages on display alerting to problems including no cap, over range and under range readings, and light source error.
GLP
• Records the date of the last user calibration performed.
Indexing indent on meter
• The meter has an indent that allows for a lock and key fit with the cuvette cap.
• This ensures that the cuvette is indexed consistently (same position) in order to maintain the same path length for accurate results.
Battery status indicator
The instrument has a battery level indicator to display reaming life as follows:
• 3 lines for 100 % capacity
• 2 lines for 66 % capacity
• 1 line for 33 % capacity
• Battery icon blinks when the capacity is less than 10 %.
ABSORBANCE SPECIFICATIONS
Light Source : | tungsten lamp |
---|---|
Light Detector : | silicon photocell with narrow band interference filter @ 575 nm |
ADDITIONAL SPECIFICATIONS
Range : | 0.000 to 1.000 mg/L (ppm) |
---|---|
Resolution : | 0.001 mg/L (ppm) |
Accuracy @ 25°C (77°F) : | ±0.010 mg/L ±7% of reading |
Method : | adaptation of the 1-(2-pyridylazo)-2-naphtol PAN method |
Auto-off : | after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder |
Environment : | 0 to 50°C (32 to 122°F); RH max 95% non-condensing |
Weight : | 360g (12.7oz.) |
Battery Type : | 9V battery |
Dimensions : | 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”) |
Ordering Information : | HI96740 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual.
CAL Check™ standards and testing reagents sold separately. |
Warranty : | 2 years |
Order now