Titanium Body pH Electrode for Boilers and Cooling Towers with BNC + Phono Connector
HI72911B
Description
The HI72911B is a titanium body, polymer filled, double reference junction pH electrode with a BNC + phono connector. This flat tip sensing portion of this electrode is made with general purpose glass. The HI72911B has a PTFE type junction along with a built-in temperature sensor. This design consideration is ideal for measuring the pH in boilers and cooling towers. The HI72911B is designed for use with Hanna’s HI98191 pH meter.
• Built-In Temperature Sensor for Automatic Temperature Compensation
• Large Annular PTFE Junction with Increased Surface Area for the Reference Cell
• Unbreakable Titanium Body
The HI72911B uses general purpose (GP) glass, flat tip bulb, titanium body, and PTFE junction with polymer electrolyte.
General Purpose Glass Formulation
• The measurement of pH at very high temperatures is detrimental to the sensitive glass bulb and will shorten the life of it.
• A pH electrode with general purpose (GP) glass will have a resistance of 100 megaohms at 25°C while the resistance of LT glass is around 50 megaohms at 25°C.
• As the temperature of the glass decreases in the sample, the resistance of the LT glass will approach that of GP glass.
• If using GP glass, the resistance would increase above the optimum range, resulting in increased impedance and ultimately affecting the measurement.
• The HI72911D is suitable to use with samples that measure from 0 to 80°C.
Flat Tip Bulb
• The recessed flat tip bulb of the HI72911B is easy to clean and prevents solids in solution from collecting on the sensor.
• Other tip shapes include conic for penetration and spheric for aqueous measurements.
Titanium Body
• A titanium body increases immunity to electrostatic and magnetic fields.
• It also allows for strong corrosion resistance, even in seawater.
PTFE Junction
• Porous polytetrafluoroethylene (PTFE) is a hydrophobic material that is available with different porosities.
• This type of junction is often used on electrodes with polymer electrolytes. Because of its chemical advantages, PTFE is widely used in industrial applications.
BNC + RCA Connector
• The HI72911B uses a BNC + RCA connector.
• This type of connector is universal in that it can be used on Hanna pH meters that have a female BNC connector for pH and RCA input for temperature.
• Other type of connectors include DIN, screw type, T-type, and 3.5mm to name a few.
• These types of connectors tend to be proprietary for a particular type of meter and are not interchangeable.
SPECIFICATIONS
Body Material : | titanium |
---|---|
Reference : | double, Ag/AgCl |
Junction / Flow Rate : | PTFE |
Electrolyte : | polymer |
Range : | pH: 0 to 13 |
Max Pressure : | 3.0 bar |
Tip Shape : | flat |
Diameter : | 14 mm |
Body Length / Overall Length : | 110 mm / 150.5 mm |
Recommended Operating
Temperature : |
0 to 80°C (32 to 176°F) – GP |
Temperature Sensor : | yes |
Matching Pin : | no |
Amplifier : | no |
Digital : | no |
Cable : | coaxial; 1 m (3.3’) |
Connection : | BNC + phono |
Applications : | boilers, cooling towers |
Warranty : | 6 months |
Order now
To make your order, please contact us at info@polyscientific.com.my or call us at +6 06 3350 690