Titanium Body pH Electrode for Plating Baths with DIN Connector
HI62911D
Description
The HI62911D is a titanium body, polymer filled, double reference junction pH electrode with a DIN connector. The flat tip sensing portion of this electrode is made with general purpose glass. The HI62911D also has a PTFE type junction along with a built-in temperature sensor and amplifier within a single probe design. This design consideration is ideal for measuring the pH in plating baths. The HI62911D is designed for use with Hanna’s HI99131 pH meter.
• Built-In Temperature Sensor
• Recessed, Flat Tip
• Titanium Body
The HI62911D uses general purpose (GP) glass, flat tip bulb, titanium body, and PTFE junction with polymer electrolyte.
General Purpose Glass Formulation
• The measurement of pH at very high temperatures is detrimental to the sensitive glass bulb and will shorten the life of it.
• A pH electrode with general purpose (GP) glass will have a resistance of 100 megaohms at 25°C while the resistance of LT glass is around 50 megaohms at 25°C.
• As the temperature of the glass decreases in the sample, the resistance of the LT glass will approach that of GP glass.
• If using GP glass, the resistance would increase above the optimum range, resulting in increased impedance and ultimately affecting the measurement.
• The HI62911D is suitable to use with samples that measure from 0 to 80°C.
Flat Tip Bulb
• The recessed flat tip bulb of the HI62911D is easy to clean and prevents solids in solution from collecting on the sensor.
• Other tip shapes include conic for penetration and spheric for aqueous measurements.
Titanium Body
• A titanium body increases immunity to electrostatic and magnetic fields.
• It also allows for strong corrosion resistance, even in seawater. Out titanium bodied electrodes’ outer casing also serves as a matching pin.
Matching Pin
• A matching pin is a differential measurement technique used to eliminate ground loops and common mode perturbations for the measurement system.
• In a system without a matching pin, electrical currents in the sample can affect the reference half-cell voltage that is connected via the liquid junction with the sample.
• In this case, the reference electrode picks up the electromagnetic fields and the measurement of the pH is altered.
• The matching pin isolates these current/magnetic fields from the reference electrode. The HI62911D titanium body serves as a matching pin for safe precise pH measurements.
PTFE Junction
• Porous polytetrafluoroethylene (PTFE) is a hydrophobic material that is available with different porosities.
• This type of junction is often used on electrodes with polymer electrolytes. Because of its chemical advantages, PTFE is widely used in industrial applications.
DIN Connector
• The HI62911D uses a DIN connector.
• This type of connector is generally proprietary to the meters they are supplied with and may not be interchangeable.
• The HI62911D is designed for use with Hanna’s HI99131 pH meter.
• Other type of connectors include BNC, screw type, T-type, and 3.5mm to name a few.
SPECIFICATIONS
Body Material : | titanium body working as matching pin |
---|---|
Reference : | double, Ag/AgCl |
Junction / Flow Rate : | PTFE |
Electrolyte : | polymer |
Range : | pH: 0 to 13 |
Max Pressure : | 3.0 bar |
Tip Shape : | flat |
Diameter : | 14 mm |
Body Length / Overall Length : | 120 mm / 160 mm |
Recommended Operating
Temperature : |
0 to 80°C (32 to 176°F) – GP |
Temperature Sensor : | yes |
Matching Pin : | yes |
Amplifier : | yes |
Digital : | no |
Cable : | 7-pole; 1 m (3.3′) |
Connection : | DIN |
Applications : | plating |
Warranty : | 6 months |
Order now
To make your order, please contact us at info@polyscientific.com.my or call us at +6 06 3350 690