HI4003

Cadmium Half-Cell Ion Selective Electrode (ISE)

Description

The HI4003 is a solid state, half-cell ion selective electrode (ISE) for the determination of cadmium (Cd2+) in solution. The mixed cadmium sulfide/silver sulfide sensing pellet produces a potential change due to the cadmium ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable epoxy body. The HI4003 is ideal for a variety of applications in electroplating, battery construction, and in the laboratory.

•  Detection from 0.01 to 11,200 mg/L Cd2+
•  For use with HI5315 reference half-cell
•  Solid state sensor

The HI4003 cadmium ion selective electrode is a potentiometric half-cell sensor housed within an epoxy body. The internal elements are isolated from the sample by a compressed cadmium sulfide/silver sulfide sensing membrane. This solid state sensor requires a separate reference electrode to complete the measurement system.

Hanna’s solid state half-cell ISEs are designed for use with the HI5315 reference half-cell. The reference half-cell features a cone style junction. The skirted cone geometry of the reference forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.

For the HI4003 to measure cadmium accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the HI4000-00  ionic strength adjuster (ISA) solution. The ISA ensures that the activity coefficient is constant so the concentration of free cadmium ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI4003 cadmium solid state ISE can measure from 0.1µM (0.01 mg/L) to 0.1M (11,200 mg/L).

    Conversion Table for Cd2+

  Multiply
    moles/L (M) to ppm (mg/L) :  1.124 x 105
    ppm (mg/L) to M (moles/L) :  8.896 x 10-6

Features at-a-glance
BNC Connection
•  The HI4003 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.

  Durable Epoxy Body
•  The probe body of the cyanide half-cell ISE is composed of durable epoxy resin.

  Fixed Sensing Modules
•  Solid state half-cell ISEs are electrodes that detect free ions in a solution.
•  The highly insoluble, compressed sensing pellet is attached to the HI4003 cadmium half-cell by a fixed sensing module for ease of use.

     SPECIFICATIONS

    Type                          :  solid state; half-cell
    Measurement Range  :  Cadmium (Cd2+) 0.1M to 1•10-7M, 11,200 to 0.01 mg/L (ppm)
    Optimum pH Range   :  pH 2 to 12
    Temperature Range   :  0 to 80 °C
    Approximate Slope    :  +28 mV
    Diameter                    :  12 mm
    Overall Length           :  120 mm
    Body Material            :  epoxy
    Cable                         :  coaxial; 1 m (3.3’)
    Connection                :  BNC
    Warranty                    :  6 months