Bromide Half-Cell Ion Selective Electrode (ISE)
HI4002
Description
The HI4002 is a solid state, half-cell ion selective electrode (ISE) for the determination of bromide (Br-) in solution. The silver bromide sensing pellet produces a potential change due to the bromide ion exchange between the inorganic pellet membrane and the sample. The internal sensing elements are housed within a durable epoxy body. The HI4002 is ideal for a variety of applications in the pharmaceutical industry, laboratories, drinking water and wastewater treatment, food and beverage production, and in printing processes.
• Detection from 0.08 to 79,910 mg/L Br-
• Requires reference half-cell
• Solid state sensor
The HI4002 bromide ion selective electrode is a potentiometric half-cell sensor housed within an epoxy body. The internal elements are isolated from the sample by a compressed silver bromide sensing membrane. This solid state sensor requires a separate reference electrode to complete the measurement system.
Hanna’s solid state half-cell ISEs are designed for use with the HI5315 reference half-cell. The reference half-cell features a cone style junction. The skirted cone geometry of the reference forms the liquid junction with the test solution at the tip of the junction cone, producing a highly stable reference potential and consistent flow rates of electrolyte into the solution. A simple press of the top portion of the ISE makes it easy to empty the electrolyte fill solution and replace when necessary.
For the HI4002 to measure bromide accurately, it is important that the ionic strength of the standards and sample be adjusted to a high and constant value with the halide ionic strength adjuster (ISA) solution. The halide ISA ensures that the activity coefficient is constant so the concentration of free bromide ions can be measured. Measuring all solutions with a constant ionic strength reduces the margin of error between measurements. The HI4002 bromide solid state ISE can measure from 1µM (0.08 mg/L) to 1M (79,910 mg/L).
Conversion Table for Br–
Multiply |
|
moles/L (M) to ppm (mg/L) : |
7.99 x 104 |
ppm (mg/L) to M (moles/L) : |
1.252 x 10-5 |
Features at-a-glance
BNC Connection
• The HI4002 has a universal BNC connector for easy connection to any benchtop meter with a female BNC probe input.
Durable Epoxy Body
• The probe body of the cyanide half-cell ISE is composed of durable epoxy resin.
Fixed Sensing Modules
• Solid state half-cell ISEs are electrodes that detect free ions in a solution.
• The highly insoluble, compressed sensing pellet is attached to the HI4002 bromide half-cell by a fixed sensing module for ease of use.
SPECIFICATIONS
Type : | solid state; half-cell |
---|---|
Measurement Range : | Bromide (Br–) 1.0M to 1•10-6M, 79,910 to 0.08 mg/L (ppm) |
Optimum pH Range : | pH 2 to 12.5 |
Temperature Range : | 0 to 80 °C |
Approximate Slope : | -56 mV |
Diameter : | 12 mm |
Overall Length : | 120 mm |
Body Material : | epoxy |
Cable : | coaxial; 1 m (3.3’) |
Connection : | BNC |
Warranty : | 6 months |
Order now
To make your order, please contact us at info@polyscientific.com.my or call us at +6 06 3350 690