Dissolved Oxygen Chemical Test Kit
HI3810
Description
The HI3810 is a chemical test kit that measures dissolved oxygen by modified Winkler titration. The HI3810 is supplied complete with all of the reagents and equipment necessary to perform approximately 110 tests. The HI3810 is a titration-based chemical test kit that determines the dissolved oxygen concentration within the 0 to 10 mg/L Ol2 range.
• All reagents marked with expiration date and lot number for traceability
• Manual titration performed with starch indicator
• Pre-made reagents for ease of use
Features at-a-glance
Complete setup
• All required materials are included with the test kit, such as the glass stoppered bottle, indicator and reagent bottles, and calibrated syringe.
High resolution
• Readings from 0 to 10 mg/L are determined to 0.1 mg/L resolution.
Replacement reagents available
• There is no need to buy a new kit when reagents are exhausted. The HI3810-100 can be ordered to replace the reagents supplied with the kit.
Significance of Use
• The concentration of dissolved oxygen in water is extremely important in nature as well in man’s environment. In oceans, lakes, rivers, and other surface water bodies, dissolved oxygen is essential to the growth and development of aquatic life.
Without oxygen, water can become toxic due to the anaerobic decaying of organic matter. In man’s environment, water must contain at least 2 mg/L of oxygen to protect water pipes from corrosion. However, boiler system water, in many cases,
cannot contain greater than 10 mg/L oxygen.
A modified Winkler method is used in the HI3810 test kit. Manganous ions react with oxygen in the presence of potassium hydroxide to form a manganese oxide precipitate (Step 1). An azide is present to prevent any nitrite ions from interfering with the test
With addition of acid, manganese oxide hydroxide oxidizes the iodide to iodine (Step 2). Since the amount of iodine generated is equivalent to the oxygen in the sample, the concentration of iodine is calculated by titration of thiosulfate ions that reduce the
iodine back to iodide ions (Step 3).
• Step 1: 2Mn2+ + O2 + 4OH– → 2MnO(OH)2
• Step 2: MnO(OH)2 + 2I– + 4H+ → Mn2+ + I2 + 3H2O
• Step 3: I2 + 2S2O32– + S4O62-
SPECIFICATIONS
Range : | 0.0 to 10.0 mg/L O₂ |
---|---|
Resolution : | 0.1 m/L |
Method : | modified Winkler |
Number of Tests : | 110 |
CTK Type : | Titration |
Ordering Information : | HI3810 test kit comes with 30 mL manganous sulfate solution, 30 mL alkali-azide reagent, 60 mL sulfuric acid solution (2), 10 mL starch indicator, 120 mL titrant solution,
glass bottle with stopper, 10 mL calibrated vessel, and calibrated syringe with tip. |
Order now
To make your order, please contact us at info@polyscientific.com.my or call us at +6 06 3350 690