edge® Multiparameter EC/TDS/Salinity Meter
HI2030-01
Description
edge’s groundbreaking design is the culmination of Hanna’s vision, design capabilities, integrated production and world class R&D. The edge meter is only 0.5” thick yet rich in features to accommodate the needs of a vast amount of customers. For those that prefer very simplistic operation there is a basic mode operation with simplified menu and options while for those who require advanced features there is the full featured standard operating mode. The edge HI2030 EC/TDS/Salinity kit can can be upgraded at any time with additional probes to measure pH or Dissolved Oxygen.
• edge features a 5.5” LCD display that you can clearly view from over 5 meters. The large display and it’s wide 150° viewing angle provide one of the easiest to read LCD’s
in the industry.
• edge features a capacitive touch keypad that gives a distinctive, modern look. Since the keypad is part of the screen, your buttons can never get clogged with sample residue.
• edge features clear, full text guides displayed on the bottom of the screen. There is no need to decipher scrambled abbreviations or symbols; these helpful messages guide
you through every process quickly and easily.
pH SPECIFICATIONS
pH Range : | -2.000 to 16.000 pH, -2.00 to 16.00 pH, ±1000 mV |
---|---|
pH Resolution : | 0.001 pH, 0.01 pH, 0.1 mV |
pH Accuracy (@25ºC/77ºF) : | ±0.2 mV, ±0.01 pH, ±0.002 pH |
pH Calibration : | up to 5 points |
pH Calibration Buffers : | 1.68, 10.01, 12.45, 4.01, 6.86, 7.01, 9.18, two custom buffers |
pH Temperature Compensation : | ATC: -5.0 to 100.0ºC; 23.0 to 212.0°F* |
EC SPECIFICATIONS
EC Range : | 0.00 to 29.99 μS/cm, 30.0 to 299.9 μS/cm, 300 to 2999 μS/cm, 3.00 to 29.99 mS/cm, 30.0 to 200.0 mS/cm, up to 500.0 mS/cm (absolute EC)** |
---|---|
EC Resolution : | 0.01 μS/cm, 0.1 μS/cm, 1 μS/cm, 0.01 mS/cm, 0.1 mS/cm |
EC Accuracy (@25ºC/77ºF) : | ±1% of reading (±0.05 μS/cm or 1 digit, whichever is greater) |
EC Calibration : | 1 point offset calibration (0.00 μS/cm in air), 1 point slope calibration in EC standard 84 μS/cm, 1413 μS/cm, 5.00 mS/cm,
12.88 mS/cm, 80.0 mS/cm and 111.8 mS/cm |
TDS SPECIFICATIONS
TDS Range : | 0.00 to 14.99 ppm (mg/L), 15.0 to 149.9 ppm (mg/L), 150 to 1499 ppm (mg/L), 1.50 to 14.99 g/L, 15.0 to 100.0 g/L,
up to 400.0 g/L (absolute TDS)**, with 0.80 conversion factor |
---|---|
TDS Resolution : | 0.01 ppm, 0.1 ppm, 1 ppm, 0.01 g/L, 0.1 g/L |
TDS Accuracy (@25ºC/77ºF) : | ±1% of reading (±0.03 ppm or 1 digit, whichever is greater) |
TDS Calibration : | through EC calibration |
TDS Conversion Factor : | 0.40 to 0.80 |
SALINITY SPECIFICATIONS
Salinity Range : | 0.0 to 400.0% NaCl, 0.01 to 42.00 PSU, 0.0 to 80.0 g/L* |
---|---|
Salinity Resolution : | 0.1% NaCl, 0.01 PSU, 0.01 g/L |
Salinity Accuracy (@25ºC/77ºF) : | ±1% of reading |
Salinity Calibration : | 1 point with HI 7037L 100% NaCl sea water standard (other scales through EC calibration) |
DO SPECIFICATIONS
DO Range : | 0.00 to 45.00 ppm (mg/L), 0.0 up to 300.0% saturation |
---|---|
DO Resolution : | 0.01 ppm, 0.1% saturation |
DO Accuracy : | ±1 digit, ±1.5% of reading |
DO Calibration : | one or two points at 0% (HI 7040 solution) and 100% (in air) |
DO Temperature Compensation : | ATC: 0 to 50 ºC; 32.0 to 122.0 °F |
Salinity Compensation : | 0 to 40 g/L (with 1 g/L resolution) |
DO Altitude Compensation : | -500 to 4000 m (with 100 m resolution) |
TEMPERATURE SPECIFICATIONS
Temperature Range : | -20.0 to 120.0 ºC, -4.0 to 248.0 °F |
---|---|
Temperature Resolution : | 0.1 °C, 0.1 °F |
Temperature Accuracy : | ±0.5 °C, ±1.0 °F |