HI96741

Total Hardness and Low Range Iron Portable Photometer

Description

The HI96741 Total Hardness and Low Range Iron Portable Photometer combines accuracy and ease of use in an ergonomic, portable design. A user can accurately determine these two parameters using ready made reagents. The HI96741 offers many advanced features including the exclusive CAL Check™ function used for performance verification and calibration of the meter.

•  Automatic Shut Off
•  Built-in Timer
•  CAL Check™ Validation and Calibration

The HI96741 portable photometer is for the measurement of total hardness and low range iron. Hanna’s portable photometers feature an advanced optical system; the combination of a special tungsten lamp, a narrow band interference filter, and silicon photodetector ensure accurate photometric readings every time. The Hanna exclusive CAL Check feature utilizes ready-made, NIST traceable standards to verify both meter validation and calibration. The exclusive cuvette locking system ensures that the cuvette is inserted into the measurement cell in the same position every time to maintain a consistent path length.

   ABSORBANCE SPECIFICATIONS

  Light Source   : tungsten lamp
  Light Detector : silicon photocell with narrow band interference filter @ 525 nm

   IRON SPECIFICATIONS

  Iron Range        : 0 to 1.60 mg/L (ppm)
  Iron Resolution : 0.01 mg/L
  Iron Accuracy   : ±0.01 mg/L ±8% of reading
  Iron Method      : adaptation of the TPTZ method

   CALCIUM SPECIFICATIONS

  Calcium Hardness Range        : 0.00 to 2.70 mg/L (ppm) as CaCO3
  Calcium Hardness Resolution : 0.01 mg/L
  Calcium Hardness Accuracy   : ±0.11 mg/L ±5% of reading
  Calcium Hardness Method      : adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, colorimetric method

   TOTAL HARDNESS SPECIFICATIONS

  Hardness Range        : 0.00 to 4.70 mg/L (ppm) as CaCO3
  Hardness Resolution : 0.01 mg/L
  Hardness Accuracy   : ±0.11 mg/L ±5% of reading
  Hardness Method      : adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, colorimetric method

   MAGNESIUM SPECIFICATIONS

  Magnesium Hardness Range        : 0.00 to 2.00 mg/L (ppm) as CaCO3
  Magnesium Hardness Resolution : 0.01 mg/L
  Magnesium Hardness Accuracy   : ±0.11 mg/L ±5% of reading
  Magnesium Hardness Method      : adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th Edition, colorimetric method

   ADDITIONAL SPECIFICATIONS

  Range                            : 0 to 300 μg/L (ppb)
  Resolution                     : 1 µg/L
  Accuracy @ 25°C (77°F) : ±10 μg/L ±3% of reading
  Method                          : adaptation of the 1-(2-pyridylazo)-2-naphtol PAN method
  Auto-off                         : after ten minutes of non-use in measurement mode; after one hour of non-use in calibration mode; with last reading reminder
  Environment                 : 0 to 50°C (32 to 122°F); RH max 95% non-condensing
  Weight                          : 360g (12.7oz.)
  Battery Type                  : 9V battery
  Dimensions                   : 193 x 104 x 69 mm (7.6 x 4.1 x 2.7”)
  Ordering Information     : HI96748 is supplied with sample cuvettes (2) with caps, 9V battery, instrument quality certificate and instruction manual.

CAL Check™ standards and testing reagents sold separately.

  Warranty                       : 2 years

Order now

To make your order, please contact us at  info@polyscientific.com.my or call us at +6 06 3350 690

Go to Top