For scientists and professionals who need precise water content determination from 1 ppm to 5%, the HI904 is a reliable Coulometric Karl Fischer titrator that gives results in minutes. Unlike volumetric Karl Fischer titrators, the HI904 features an precision titrant generation system and a fully sealed titration vessel for unmatched accuracy.
Range : |
1 ppm to 5% |
Resolution : |
0.1ppm to 0.0001% |
Result Units : |
%, ppm, ppt, mg/g, μg/g, mg, μg, mg/mL, μg/mL, mg Br/100g, g Br/100g, mg Br, g Br |
Sample Types : |
liquid or solid with external dissolution / extraction |
Pre-Titration Conditioning : |
automatic |
Background Drift Correction : |
automatic or user-selectable value |
Endpoint Criteria : |
fixed mV persistence, relative drift stop or absolute drift stop |
Dosing : |
dynamic |
Result Statistic : |
mean, standard deviation |
Generator Configuration : |
diaphragm or diaphragm-less |
Generator Current Control : |
automatic or fixed (400 mA) |
Generator Type Detection : |
automatic |
Titration Vessel : |
operating volume between 100 – 200 mL |
Solvent Handling System : |
sealed system with integrated diaphragm air pump and beaker adapter |
Sensor Type : |
HI76330 dual platinum pin polarization electrode |
Sensor Connection : |
BNC |
Polarization Current : |
1, 2, 5, or 10 μA |
Voltage Range : |
2 mV to 1100 mV |
Voltage Resolution : |
0.1 mV |
Accuracy (@25ºC/77ºF) : |
±0.1% |
Programmable Stirrer : |
magnetic, optically regulated, digital stirrer; 200 to 2000 rpm; resolution 100 rpm |
Display : |
5.7” (320 x 240 pixel) backlit color LCD |
Methods : |
up to 100 (standard and user) methods |
Logging Data : |
up to 100 complete titration reports and drift rate reports can be stored |
Peripherals (Rear) : |
connections for VGA display, PC-keyboard, parallel printer, USB device input, RS232 |
GLP Conformity : |
Good Laboratory Practice and instrument data storage and printing |
Languages : |
English, Portuguese. Spanish, and French |
Enclosure Material : |
ABS plastic and steel |
Keypad : |
polycarbonate |
Operating Environment : |
10 to 40°C (50 to 104°F), up to 95% RH |
Storage Environment : |
-20 to 70°C (-4 to 158°F), up to 95% RH |
Dimensions : |
390 x 350 x 380 mm (15.3 x 13.8 x 14.9”) |
Weight : |
approximately 10 kg (22 lbs.) |