HI83141

Analog pH/mV/°C Meter with HI1230B Electrode

Description

The HI83141 is a portable pH/mV meter designed to be accurate, reliable and easy to use. The HI83141 uses the HI1230B pH electrode and HI7669AW temperature probe using separate connections.

•  Automatic Temperature Compensation
•  Compact, heavy-duty, and waterproof protected casing
•  Two-point Calibration

Manual calibration is performed at one or two points by adjusting the trimmers on the front panel. Capable of measuring pH/mV and °C, this meter are great for field work, providing one meter for multiple uses. This instrument is ideal for applications that require a custom calibration point. Manual calibration can be extremely useful in order to achieve better accuracy.

This meter can also be used for ORP measurements with the optional probes below: HI3131B

   pH SPECIFICATIONS

  pH Range                  : 0.00 to 14.00 pH
  pH Resolution           : 0.01 pH
  pH Accuracy             : ±0.01 pH
  pH Calibration          : manual, two-point, via trimmers
  pH Temperature

Compensation          :

 

automatic, 0 to 70°C (32 to 158 °F)

  pH Input impedence  : 1012 Ohm

   mV SPECIFICATIONS

  mV Range          : ±1999 mV
  mV Resolution   : 1 mV
  mV Accuracy     : ±1 mV

   METER TEMPERATURE SPECIFICATIONS

  Temperature Range       : 0.0 to 100.0°C
  Temperature Resolution : 0.1°C
  Temperature Accuracy   : ±0.4°C

   GENERAL METER SPECIFICATIONS

  pH Electrode : HI1230B PEI body pH electrode with BNC connector and 1 m (3.3’) cable (included)
  Temperature

Probe :

 

HI7669AW stainless steel temperature probe, BNC connector (included)

  Slope/Offset

Calibration :

 

from 80 to 110%/±1 pH

  Battery Type /Life :  

9V / approximately 100 hours of continuous use

  Environment  : 0 to 50°C (32 to 122°F); RH max 100%
  Dimensions    : 145 x 80 x 36 mm (5.7 x 3.1 x 1.4’’)
  Weight              : 230 g (8.1 oz.)
  Ordering

Information :

HI83141 is supplied with HI1230B pH electrode and HI7669AW temperature probe, HI70004 pH 4.01 buffer solution sachet,

HI70007 pH 7.01 buffer solution sachet, HI700601 electrode cleaning solution sachets (2), calibration screwdriver, battery, protective case and instructions.

Order now

To make your order, please contact us at  info@polyscientific.com.my or call us at +6 06 3350 690

Go to Top