HI8033
Description
The HI8033 is a durable, portable EC/TDS meter designed for use in a wide variety of applications. It is supplied with a four-ring conductivity probe to allow for three ranges of EC measurement and one range of TDS measurement. The temperature compensation coefficient (β) is fixed at 2%/°C and all conductivity readings are manually adjusted for temperature, assuring users of consistent and accurate measurements.
• Four-ring probe
• Manual Temperature Compensation
• Three measuring ranges
The portable HI8033 EC/TDS meter is suitable for use in a variety of applications. It offers three EC measurement ranges from 0.0 µS/cm to 19.99 mS/cm and a TDS measurement range from 0 to 19990 mg/L (ppm) with a ±1% FS accuracy. The HI76301W four-ring conductivity probe that is supplied with the meter allows for a wide range of measurements with a single sensor. The four-ring technology also eliminates the polarization effect that is common with standard two pole versions. The HI8033 offers manual temperature compensation from 0 to 50°C (32 to 122°F) and a fixed β at 2%/°C.
Features/Benefits:
Fixed Temperature Coefficient
• As the temperature of a solution increases or decreases, the conductance becomes higher or lower, respectively.
• The temperature compensation coefficient, also known as β, is the factor that adjusts for the temperature of a solution to ensure an accurate reading is achieved.
• The HI8033 is programmed with a fixed β at 2%/°C; this factor corrects the conductivity reading by 2% for each degree Celsius change in the sample.
Four-ring Probe
• The HI76301W four-ring probe that comes with the HI8033 offers a versatile and accurate solution for conductivity readings.
• Four-ring technology allows for a larger range of measurement within a single probe, whereas two probe technology is somewhat limited in the range in which it can measure.
• The probe comes with a protective PVC sleeve that assures durability when taking measurements outdoors.
Manual Temperature Compensation
• Since temperature has such a dramatic effect on conductivity readings, having a meter that offers temperature compensated readings is invaluable.
• The HI8033 allows users to adjust the temperature of the measurement solution through a knob on the face of the meter.
• The adjustment accounts for the effects of temperature on a sample’s conductivity reading in the range of 0 to 50°C (32 to 122°F).
One-point Calibration
• The HI8033 can be calibrated at one point in a standard conductivity solution.
• The calibration knob located on the face of the meter is easily adjusted to the correct calibration standard.
Three Measurement Ranges
• HI8033 offers three conductivity measurement ranges and one TDS measurement range.
• Each range can be selected by a dial on the face of the meter, allowing users to easily switch between ranges when necessary.
• The meter is programmed to let the user know when their current reading is out of range, and a new range should then be selected.
SPECIFICATIONS
EC Range : | 0.0 to 199.9 μS/cm; 0 to 1999 μS/cm; 0.00 to 19.99 mS/cm |
---|---|
EC Resolution : | 0.1 μS/cm; 1 μS/cm; 0.01 mS/cm |
TDS Range : | 0 to 19990 mg/L (ppm) |
TDS Resolution : | 10 mg/L (ppm) |
EC/TDS Accuracy : | ±1% FS (excluding probe error) |
Temperature Compensation : | manual, 0 to 50°C (32 to 122°F) with β = 2%/°C |
Probe : | HI76301W conductivity probe with 1 m (3.3’) cable (included) |
Calibration : | manual, one-point |
Battery Type / Life : | 9V / approximately 100 hours of continuous use |
Environment : | 0 to 50°C (32 to 122°F); RH max 100% |
Dimensions : | 185 x 82 x 47 mm (7.3 x 3.2 x 1.9’’) |
Weight : | 270 g (9.5 oz.) |
Ordering Information : | HI8033 is supplied with HI76301W conductivity probe, battery, and instructions. |
Warranty : | 2 years (probe 6 months) |